Restenosis in coronary bare metal stents. Importance of time to follow-up: a comparison of coronary angiograms 6 months and 4 years after implantation

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Objectives. Angiographic late lumen loss measured 6 to 9 month after bare metal stent implantation in the coronary arteries is a validated restenosis parameter. Design. We performed a second angiographic follow-up after 4 years in event free survivors from the DANSTENT trial cohort. Results. Quantitative comparison of paired coronary angiograms at 6 months and 4 years showed a reduction of late loss from 0.68+/-0.52mm to 0.42 (+/-0.52) (mean difference 0.26 (0.17 to 0.36), p<0.0001). Minimal instent lumen diameter had increased from 2.39+/-0.62mm to 2.64+/-0.56mm (mean difference: -0.24mm, 95% confidence interval: -0.34mm to -0.14mm, p<0.0001). Instent diameter stenosis decreased from 24.8+/-14.2% to 18.6+/-9.3% (mean difference 6.16%, 95% confidence interval: 2.82 to 9.48%, p=0.0006). This observed spontaneous decrease of instent restenosis corresponds to a 19% increase of minimal cross-sectional vessel area and a 39% reduction of the binary restenosis rate over time. Conclusions. Instent late lumen loss in bare metal stents decreases spontaneously over time. Maturation of early hyperplastic tissue reaction after stent implantation with subsequent thinning of fibrotic tissue might explain this phenomenon
Udgivelsesdato: 2008/11/23
OriginalsprogEngelsk
TidsskriftScandinavian Cardiovascular Journal
Vol/bind43
Udgave nummer2
Sider (fra-til)87-93
Antal sider6
ISSN1401-7431
DOI
StatusUdgivet - 2008

ID: 10905026