Accuracy of continuous glucose monitoring during exercise-related hypoglycemia in individuals with type 1 diabetes
Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Dokumenter
- Fulltext
Forlagets udgivne version, 1,99 MB, PDF-dokument
Background: Hypoglycemia is common in individuals with type 1 diabetes, especially during exercise. We investigated the accuracy of two different continuous glucose monitoring systems during exercise-related hypoglycemia in an experimental setting.
Materials and methods: Fifteen individuals with type 1 diabetes participated in two separate euglycemic-hypoglycemic clamp days (Clamp-exercise and Clamp-rest) including five phases: 1) baseline euglycemia, 2) plasma glucose (PG) decline ± exercise, 3) 15-minute hypoglycemia ± exercise, 4) 45-minute hypoglycemia, and 5) recovery euglycemia. Interstitial PG levels were measured every five minutes, using Dexcom G6 (DG6) and FreeStyle Libre 1 (FSL1). Yellow Springs Instruments 2900 was used as PG reference method, enabling mean absolute relative difference (MARD) assessment for each phase and Clarke error grid analysis for each day.
Results: Exercise had a negative effect on FSL1 accuracy in phase 2 and 3 compared to rest (ΔMARD = +5.3 percentage points [(95% CI): 1.6, 9.1] and +13.5 percentage points [6.4, 20.5], respectively). In contrast, exercise had a positive effect on DG6 accuracy during phase 2 and 4 compared to rest (ΔMARD = -6.2 percentage points [-11.2, -1.2] and -8.4 percentage points [-12.4, -4.3], respectively). Clarke error grid analysis showed a decrease in clinically acceptable treatment decisions during Clamp-exercise for FSL1 while a contrary increase was observed for DG6.
Conclusion: Physical exercise had clinically relevant impact on the accuracy of the investigated continuous glucose monitoring systems and their ability to accurately detect hypoglycemia.
Materials and methods: Fifteen individuals with type 1 diabetes participated in two separate euglycemic-hypoglycemic clamp days (Clamp-exercise and Clamp-rest) including five phases: 1) baseline euglycemia, 2) plasma glucose (PG) decline ± exercise, 3) 15-minute hypoglycemia ± exercise, 4) 45-minute hypoglycemia, and 5) recovery euglycemia. Interstitial PG levels were measured every five minutes, using Dexcom G6 (DG6) and FreeStyle Libre 1 (FSL1). Yellow Springs Instruments 2900 was used as PG reference method, enabling mean absolute relative difference (MARD) assessment for each phase and Clarke error grid analysis for each day.
Results: Exercise had a negative effect on FSL1 accuracy in phase 2 and 3 compared to rest (ΔMARD = +5.3 percentage points [(95% CI): 1.6, 9.1] and +13.5 percentage points [6.4, 20.5], respectively). In contrast, exercise had a positive effect on DG6 accuracy during phase 2 and 4 compared to rest (ΔMARD = -6.2 percentage points [-11.2, -1.2] and -8.4 percentage points [-12.4, -4.3], respectively). Clarke error grid analysis showed a decrease in clinically acceptable treatment decisions during Clamp-exercise for FSL1 while a contrary increase was observed for DG6.
Conclusion: Physical exercise had clinically relevant impact on the accuracy of the investigated continuous glucose monitoring systems and their ability to accurately detect hypoglycemia.