Continuous direct tissue oxygen tension measurement. A new application for an intravascular oxygen sensor

Research output: Contribution to journalJournal articleResearchpeer-review

An oxygen tension measurement equipment, developed for continuous intravascular monitoring was tested as a direct tissue oxygen tension meter. The sterile disposable oxygen probe consists of a Clark type polaro-graphic bipolar electrode placed in a 5 cm long flexible polyethylene catheter (OD 0.55 mm). The probe is easily placed, e.g. in subcutaneous tissue or muscle, with the aid of an introduction needle. In vitro measurements as well as subcutaneous and intramuscular readings were performed. The animals were subjected to various experimental conditions such as different inspiratory oxygen fractions, hypovolemia, and circulatory impairment. Results were comparable to those obtained with previously used and well characterized tissue oxygen tension measurement equipment. The clinical applicability of the oxygen sensor for tissue oxygen measurement was assessed in one patient by 40 hours' surveillance in a free myocutaneous flap. The procedure was simple and implied minimal discomfort for the patient.

Original languageEnglish
JournalAnnales Chirurgiae et Gynaecologiae. Supplementum
Volume80
Issue number1
Pages (from-to)8-13
Number of pages6
ISSN0355-9521
Publication statusPublished - 1991

    Research areas

  • Animals, Electrodes, Humans, In Vitro Techniques, Monitoring, Physiologic/instrumentation, Oxygen/analysis, Swine