Transit-time flow measurement as a predictor of coronary bypass graft failure at one year angiographic follow-up

Research output: Contribution to journalJournal articleResearchpeer-review

BACKGROUND: Transit-time flow measurement (TTFM) is a commonly used intraoperative method for evaluation of coronary artery bypass graft (CABG) anastomoses. This study was undertaken to determine whether TTFM can also be used to predict graft patency at one year postsurgery.

METHODS: Three hundred forty-five CABG patients with intraoperative graft flow measurements and one year angiographic follow-up were analyzed. Graft failure was defined as more than 50% stenosis including the "string sign." Logistic regression analysis was used to analyze the risk of graft failure after one year based on graft vessel type, anastomatic configuration, and coronary artery size.

RESULTS: Nine hundred eighty-two coronary anastomoses were performed of which 12% had signs of graft failure at one year angiographic follow-up. In internal mammary arteries (IMAs), analysis showed a 4% decrease in graft failure odds for every 1 mL/min increase in TTFM (OR = 0.96, CI = [0.93; 0.99], p = 0.005). ROC analysis showed good discriminative ability for TTFM alone AUC = 69.5% in IMA grafts. For single-vein grafts the decrease in graft failure odds was 2% for every 1 mL/min increase in TTFM (OR = 0.98; CI = [0.97; 1.00], p = 0.059) and AUC of 59.9%. There were no significant relationships between TTFM and graft failure in other graft types or graft configurations.

CONCLUSION: The TTFM method has good discriminative ability for assessing the risk of graft failure in certain graft types within the first year after CABG surgery and is a valuable instrument for intraoperative quality assessment of bypass grafts.

Original languageEnglish
JournalJournal of Cardiac Surgery
Volume30
Issue number1
Pages (from-to)47-52
Number of pages6
ISSN0886-0440
DOIs
Publication statusPublished - Jan 2015

ID: 134780829