Comparison of ankle-brachial index measured by an automated oscillometric apparatus with that by standard Doppler technique in vascular patients

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

OBJECTIVES: To evaluate the determination of ankle-brachial indices (ABIs) using a simple automated ankle pressure measurement device in comparison with the Doppler technique. DESIGN: ABI was measured in 61 patients (122 legs) admitted to the department of vascular surgery, Rigshospitalet. ABI was calculated twice using both the methods on both legs. MATERIALS AND METHODS: We tested the automated oscillometric blood pressure device, CASMED 740, for measuring ankle and arm blood pressure and compared it with the current gold standard, the hand-held Doppler technique, by the Bland-Altman analysis. RESULTS: Using the Doppler-derived ABI as the gold standard, the sensitivity and specificity of the oscillometric method for determining an ABI