Stol ikke altid på PACS: Uopdaget forglemmelse af guidewire ved anlæggelse af centralt venekateter
Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
A 74-year-old woman had a CVC inserted. Unfortunately, the guide wire was lost under the procedure, and despite a chest X-ray control it was not detected. The reason the guide wire was not identified was insufficient fine-tuning of the »local« PC monitor. We recommend regular fine-tuning of monitors used for X-ray control.