Test-retest repeatability and interobserver variation of healthy tissue metabolism using F-18-FDG PET/CT of the thorax among lung cancer patients

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Dokumenter

  • Fulltext

    Forlagets udgivne version, 1,28 MB, PDF-dokument

Objectives The aim of this study was to assess the test-retest repeatability and interobserver variation in healthy tissue (HT) metabolism using 2-deoxy-2-[F-18]fluoro-d-glucose (F-18-FDG) PET/computed tomography (PET/CT) of the thorax in lung cancer patients. Methods A retrospective analysis was conducted in 22 patients with non-small cell lung cancer who had two PET/CT scans of the thorax performed 3 days apart with no interval treatment. The maximum, mean and peak standardized uptake values (SUVs) in different HTs were measured by a single observer for the test-retest analysis and two observers for interobserver variation. Bland-Altman plots were used to assess the repeatability and interobserver variation. Intrasubject variability was evaluated using within-subject coefficients of variation (wCV). Results The wCV of test-retest SUVmean measurements in mediastinal blood pool, bone marrow, skeletal muscles and lungs was less than 20%. The left ventricle (LV) showed higher wCV (>60%) in all SUV parameters with wide limits of repeatability. High interobserver agreement was found with wCV of less than 10% in SUVmean of all HT, but up to 22% was noted in the LV. Conclusion HT metabolism is stable in a test-retest scenario and has high interobserver agreement. SUVmean was the most stable metric in organs with low FDG uptake and SUVpeak in HTs with moderate uptake. Test-retest measurements in LV were highly variable irrespective of the SUV parameters used for measurements.

OriginalsprogEngelsk
TidsskriftNuclear Medicine Communications
Vol/bind43
Udgave nummer5
Sider (fra-til)549-559
Antal sider11
ISSN0143-3636
DOI
StatusUdgivet - maj 2022

ID: 315261169